10+ measurement modes for almost every sensor and application:
The most user friendly data acquisition software OXYGEN will lead you to very quick and reliable measurement results.
The TRION-18xx-MULTI modules are 10x faster than any other similar modules on the market.
To provide excellent signal to noise ratio and aliasing free data, we equipped the module with:
Analog circuit design and AD-conversion at a glance. Better performance than any other 24-bit high-speed measurement system.
Voltage and current controlled sensor excitation for almost every sensor type. There is enough power to supply also high demand sensors:
INPUT CHANNELS | ||
---|---|---|
4 ch DSUB connector | TRION-1820-MULTI-4-D | 2 MS/s @ 24-bit |
TRION3-1820-MULTI-4-D | 2 MS/s @ 24-bit | |
TRION-1850-MULTI-4-D | 5 MS/s @ 18-bit, 2 MS/s @ 24-bit | |
8 ch OB LEMO connector | TRION3-1820-MULTI-8-L0B | 2 MS/s @ 24-bit |
TRION3-1820-MULTI-8-L0B | 5 MS/s @ 18-bit, 2 MS/s 24-bit | |
INPUT TYPES | RANGES | SUPPORTED SENSORS |
Voltage |
±0.1 mV to ±100 V freely programmable | – |
IEPE® |
100 mV to ±10 V freely programmable | IEPE® sensors |
Bridge |
1 to 1000 mV/V | 4-, 5-, 6-wire full bridge; 3-, 4-, 5-wire ½ bridge; 2-, 3-, 4-wire ¼ bridge; 120 / 350 / 1000 Ω |
Resistance |
10 Ω to 30 kΩ | Potentiometer, resistance temperature detection: Pt100, Pt200, Pt300, Pt500, Pt2000 (2-, 3-, 4-wire) |
Current |
±30 mA | 4 to 20 mA sensors; loop powered sensors |
CAN |
– | CAN 2.0 |
MSI |
MSI-BR-CH-x: 500 to 50000 pC MSI-BR-TH-x: various TC ranges MSI-BR-LVDT |
LVDT, RVDT, charge output and thermocouple sensors |
Voltage input accuracy ≤10 V | DC 0.1 Hz to 10 kHz 10 kHz to 50 kHz 50 kHz to 100 kHz |
±0.02 % of reading, ±0.02 % of range ±20 μV ±0.02 % of reading, ± 0.02 % of range ±20 μV ±0.1 % of reading, ± 0.02 % of range ±20 μV ±0.5 % of reading, ± 0.02 % of range ±20 μV |
Input impedance | Differential (100 MΩ or 1 MΩ), Single ended (50 MΩ) | |
Input coupling | DC / AC (highpass filter 0.16 Hz to 100 Hz freely programmable) | |
Spurious free SNR (10 V range) | 1 kS/s to 200 kS/s: 140 dB |