Multi channel TRION™ signal conditioning modules provide analog-digital conversion on each channel and are user exhangeable.
Use TRION™ modules with each measurement system of the DEWE2 series.
Name | Channels | Sampling [kS/s] / [bit] | Sensor supply | Voltage input range | Current | IEPE | Bridge [F/H/QU] | Thermocouple | Charge | Resistance, RTD | Others | ||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
TRION-1820-POWER-4 | 4 U + 4 I | 2000 / 18 | +9 V | 1000 VRMS | Exchangeable current modules (20A, 2A, 0.2A, 5V) | ||||||||
TRION-1600-dLV | 16 or 32 | 20 | 5 V, 12 V | ±5 / ±10 | 2x counter, CAN-bus, 8x DI, 4x DO, RS485, 2x alarm out | ||||||||
TRION-1603-LV-6-BNC | 6 ISO | 250 / 16 | - | ±100 | |||||||||
TRION-1603-LV-6-L1B | 6 ISO | 250 / 16 | 1-28 V | ±100 | TEDS | ||||||||
TRION-1620-LV-6-BNC | 6 ISO | 2000 / 24 | - | ±100 | |||||||||
TRION-1620-LV-6-L1B | 6 ISO | 2000 / 24 | 1-28 V | ±100 | TEDS | ||||||||
TRION-1620-ACC | 6 ISO | 2000 / 24 | 4 / 8 mA | ±100 | Tacho | ||||||||
TRION-1802-dLV | 16 or 32 | 200 / 100 | 5 V , 12 V | ±5 / ±10 | 2x counter, CAN-bus, 8x DI, 4x DO, RS485, 2x alarm out | ||||||||
TRION-1820-MULTI | 4 ISO | 2000/24 | 0-24 V | ±100 | CAN, CNT, TEDS | ||||||||
TRION-2402-MULTI | 4/8 ISO | 204 / 24 | 0-24 V | ±100 | CAN, TEDS | ||||||||
TRION-2402-dSTG | 6/8 DIFF | 204 / 24 | 0-13 V | ±10 | TEDS | ||||||||
TRION-2402-dACC | 6/8 DIFF | 204 / 24 | 0.1-24 mA | ±100 | Tacho, TEDS | ||||||||
TRION-2402-V | 4/8 ISO | 204 / 24 | ±1000 | ||||||||||
TRION-BASE | 16 DI(O) + 2 CNT | 2000 / - | 5 V, 12 V | DIO, IRIG-B DC, Counter | |||||||||
TRION-TIMING-V3 | 8 DIO + 1 CNT | 2000 / - | 5 V, 12 V | DIO, IRIG full, GPS 1Hz, Counter | |||||||||
TRION-CNT | 6 ISO | 204 / - | 5 V, 12 V | Counter | |||||||||
TRION-CAN | 2/4 ISO | 1000 kBit/s | 5 V, 12 V | CAN | |||||||||
TRION-VGPS-20-100-V3 | 8 DIO + 1 CNT | 2000 / - | 5 V, 12 V | DIO, Counter, GPS 100Hz | |||||||||
TRION-DI-48 | 48 ISO | 3000 / - | 5 V | Digital Input |
Simultaneous sampling TRION™ signal conditioner modules are the heart and soul of the DEWE2 series of data acquisition systems.
TRION™ modules combine the power of simultaneous sampling and 24 bit ADC technology with advanced DEWETRON signal conditioning – plug one convenient module right into your DEWE2 platform.
TRION™ modules provide the TIMING BACKBONE for recording many disparate data sources in perfect sync with each other. TRION™ modules are available for a wide variety of analog inputs, as well as digital I/O lines, synchronous counter/encoders, and timing and SYNC interfaces. Additionally, TRION™ modules are available for CAN-bus interfacing.
Choose your TRION™ module(s), plug them into your DEWE2 instrument, turn the system on and get to work!
TRION™ cards are automatically identified and configured within the software as soon as it is launched.
What is a Strain Gage (also referred to as Strain Gauge) and why do you need a strain gage signal conditioner?
Simply put, a strain gage (also referred to as strain gauge) is a resistor whose resistance varies with any applied force. These sensors pick up expansion as well as contraction and convert this “strain” into an electrical signal used in pressure sensors, weight measurements, force and torque measurements and materials analysis. Typically, the sensor is bonded to a surface to measure the minute dimensional changes that occur during expansion and contraction (or compression and tension)
Strain gauges are often used in devices for measuring acceleration, pressure, tension, and force in the manufacturing, process control and research industries. However, strain gages have also been connected to the human eye retina, insects, plastics and concrete – any material that is under investigation for strain.
In most instances the electrical signal is converted from the strain cannot be used in its raw form due to incomplete bridges in the strain gage, inconsistent voltage to the power bridge and output that is extremely small (less than 10 mV/V). This is where strain gage signal conditioners come into play.
Common conditioning includes: bridge completion, bridge excitation, excitation sensing, signal amplification, offset nulling, shunt calibration and linearization.
DEWETRON’s TRION™ signal conditioning module for strain gage sensors delivers separate analog to digital conversion on each channel, simultaneous sampling for analog (204.8 kS/s per channel), 24 bit resolution with anti aliasing filters. Our modules are available with up to 8 input channels and a variety of input connectors : RJ-45, DSUB-9, and LEMO.
Here are just a few more of the impressive features of DEWETRON’s TRION™ series strain gauge signal conditioner technicial specifications:
Supported Sensors:
4- or 6-wire full bridge
3- or 5-wire ½ bridge with internal completion (software programmable)
3- or 4-wire ¼ bridge with internal resistor for 120 and 350 Ω (software programmable)
4-wire full bridge with constant current excitation (piezoresistive bridge sensors) are supported.
Bridge resistance: 80 Ω to 10 kΩ @ ≤ 5 VDC excitation
Shunt calibration: Two internal shunt resistors 50 kΩ and 100 kΩ
Shunt and completion resistor accuracy: 0.05 % ±15 ppm/K
Automatic bridge balance: ±250 % of Range